Description
In the dynamic landscape of healthcare, the Datascope Passport V emerges as a pivotal innovation in patient monitoring, combining advanced technology with unparalleled ease of use. This sophisticated monitor is engineered to provide detailed and accurate monitoring of vital signs across a wide range of patient care settings. Designed for versatility, the Passport V seamlessly integrates non-invasive blood pressure (NIBP), pulse rate, SpO2, temperature, and ECG monitoring, along with optional CO2 measurement capabilities, ensuring comprehensive patient oversight.
Key Features and Benefits
- Comprehensive Monitoring Capabilities: The Passport V offers a full suite of monitoring parameters, including ECG, NIBP, SpO2, temperature, and respiratory rate, with the option to add CO2 monitoring, providing clinicians with a complete picture of patient health.
- High-Resolution Display: A large, high-resolution color display presents critical patient data clearly and concisely, enhancing readability and facilitating quick decision-making in fast-paced environments.
- Intuitive Operation: Designed with the user in mind, the Passport V features an intuitive user interface, with touch screen capabilities and customizable shortcuts, making navigation and operation as efficient as possible.
- Advanced Connectivity: With built-in networking capabilities, the Passport V ensures easy data transfer and integration with hospital information systems (HIS) and electronic medical records (EMR), streamlining workflow and improving documentation accuracy.
Specifications
Display
- Display Type: Color, anti-glare TFT display.
- Size: 12.1″ (246 x 184.4 mm.)
- Resolution: SVGA 800 × 600 pixels.
- Waveforms: 3 to 8.
ECG ( 3 and 5-Lead)
- Leads: I, II, III, aVR, aVL, aVF, and V.
- Cable Detection: Automatic Detection of 3 or 5-lead Sets.
- Display Sensitivity: 1.25, 2.5, 5, 10, 20, 40 mm/mV.
- ECG Signal Range: +/- 8 mV (peak-to-peak value.)
- Calibration Signal: 1 mV (peak-to-peak value.)
Frequency Response to Screen
- Surgical Mode: From 1 to 20 Hz.
- Monitor Mode: From 0.5 to 40 Hz.
- ST Mode: From 0.05 to 40 Hz.
- Diagnostic Mode: From 0.05 to 150 Hz.
- Defibrillation Protection: Withstand 360 Joule discharged current as per IEC 60601-2-27, 51.101.1.
- Baseline Recovery Time: <5 s after defibrillation (in monitor and surgical mode.)
- Defib. Synchronous Pulse: The time interval between the peaks of R-waves to the start of synchronous pulses is ≤35 ms.
ECG Analog Output
- Delay: 25 ms max.
- Sensitivity: 1 V/mV of input ±10%.
Heart Rate
- Range: 15–300 bpm adult, 15–350 bpm pediatric, 15–350 bpm neonatal.
- Accuracy: ±3 bpm or 3% at 30–250 bpm whichever is greater ±5% at 251–350 bpm, the greater of.
- Pacer Pulse Rejection: Rejects pulses of amplitude between ±2 mV to ±700 mV and duration 0.1 ms to 2 ms with no tail.
- Tall T Waves Rejection: Rejects all T-Waves of amplitude less than 1.2 mV, QRS 100 ms, QT interval of 350 ms, duration 180 ms.